General Information
Page 12 of 33 Safe Place Transmitters User Guide
0510-1121-J
Banding
NOTE: All banding material types must not
remain on a patient for longer than 72 hrs (3
days). If the material is kept on longer than
this specified period of time, you may
receive false alarms from that transmitter.
CAUTION: You must test all transmitters
prior to use to verify proper operation. This
includes every time that the band is
replaced. Failure to test the transmitters
before use can result in system failure,
abduction, and/or an elopement. In addition,
failure to test transmitters voids the RF
Technologies Product Warranty.
Smart Sense
To band the Smart Sense transmitter:
1. Remove the Smart Sense banding
material (with the plastic guide tabs still
attached) from the plastic backing.
WARNING: Only Smart Sense banding
material can be used to fasten Smart Sense
transmitters. Use of any other banding
material with the Smart Sense transmitters
voids the RF Technologies' Product
Warranty and could result in system failure.