CHAPTER 2 Installing Transceiver Devices
Page 38 of 64 Quick Response Premiere Wireless Call System Hardware Installation Guide
0510-1099-E
Setting the Call Stations Up For Use
To set up the Nurse Call transceiver for use
NOTE: Prior to any Call Device set up the devices must be assigned to a unit. This is what
allows the device to be seen in the Event List. Refer to Series Software Administrator Guide.
1. Plug in the desired Nurse Call Cord set into the Nurse Call transceiver.
2. Activate the Nurse Call transceiver by pressing the button on the cord set.
3. If the Nurse Call transceiver is working properly, an Assistance Required alarm event will be listed
on the Event List at the Central Server. The system senses the Nurse Call transceiver when it goes
into alarm and adds it to its list of devices.
For latching type cord sets reset the Nurse Call transceiver by pulling the button back to the UP
position. Momentary type cord sets can be reset by pressing the reset button on the side of the
device. The alarm event will automatically clear from the Event List at the Central Server. If
Enforce Joint Commission is enabled, you must clear the alarm at the Central Server or Client
computer by selecting a reason.
4. At the Central Server, update the Nurse Call transceiver information, for example, giving the Nurse
Call a name and/or assigning it to a room or unit. Refer to the “Update Devices” section in the
applicable Series Software Administrator Guide.
To set up the Push Button Emergency Call transceiver for use
1. Activate the Push Button Emergency Call transceiver by pushing the red button. When pushed, the
button remains in a recessed position indicating the device is in an alarm state.
If the Push Button Emergency Call transceiver is working properly, an Assistance Required alarm
event will be listed on the Event List at the Central Server. The system senses the Push Button
Emergency Call when it goes into alarm and adds it to its list of devices.
2. Reset the Push Button Emergency Call by pushing the red button again, the button pops out
indicating the transceiver has changes states and is now idle and ready for its next usage.
The alarm event will automatically clear from the Event List at the Central Server. If Enforce Joint
Commission is enabled, you must clear the alarm at the Central Server by selecting a reason.
3. At the Central Server, update the Push Button Emergency Call information, for example, giving the
Push Button Emergency Call a name and/or assigning it to a room or unit. Refer to the “Update
Devices” section in the applicable Series Software Administrator Guide.
To set up the Pull Cord or Pull Cord with Check In for use
1. Activate the Pull Cord by pulling the cord.
2. If the transceiver is working properly, an Assistance Required alarm event will be listed on the
3. Event List at the Central Server. The system senses the transceiver when it goes into alarm and adds
it to its list of devices.
4. Reset the Pull Cord by rotating the red HELP lever back to the UP position.
5. The alarm event will automatically clear from the Event List at the Central Server. If Enforce Joint
Commission is enabled, you must clear the white alarm at the Central Server by selecting a reason.
6. At the Central Server, update the Pull Cord information, for example, giving the Pull Cord a name
and/or assigning it to a room or unit. Refer to the “Update Devices” section in the applicable Series
Software Administrator Guide.