CHAPTER 1 Installing Hardware Components
Quick Response Premiere Wireless Call System Hardware Installation Guide Page 25 of 64
0510-1099-E
Test the 32 Chanel Controller
When an alarm is sent from the Server to the Router, the 32 Channel Controller acts as a relay
switch and turns on the Dome Light mounted outside the patient’s room. Once the 32 Channel
Controller /Dome Light setup has been configured and wired properly, you must test its
operation.
1. Position yourself where you can see the Dome Light.
2. Have someone activate a transceiver device.
3. If the 32 Channel Controller and Dome Light have been wired correctly, the Dome Light
will illuminate and an Alarm event will be listed on the Event List at the Central Server.
4. Reset the transceiver device and follow step 1 3 to test all 32 Channel Controller/Dome
Light setups.
NOTE: The Dome Light includes two indicator lights: a white indicator light for low
priority alarms and a red indicator light for high priority alarms.
Emergency CallsEmergency calls are indicated by steady illumination of the
red Dome Light associated with the zone or area from which an emergency call
has been placed.
Normal CallsNormal calls are indicated by steady illumination of the white
Dome Light associated with the zone or area from which a normal call has
been placed.
Concurrent Emergency and Normal CallsIf emergency and normal calls are
placed at the same time, the red light connected to the emergency station from
which a call was placed will illuminate at the same time as the white light
associated with the normal call.