User Manual
Table Of Contents
- Important Warnings
- Bio-Incompatibility Notice
- Compliance
- Introduction
- About this Guide
- Additional Detailed Documentation
- Contact Information
- Product Warranty
- Chapter 1
- Introduction
- Installation Checklist
- Installing Components
- Install the Central Server
- Install the Gateway
- Install Routers
- Place a Pendant into Survey Mode
- Determine Placement of Routers
- Quick Look Router
- Gateway/Router Reset Button
- LED Sequence
- Gateway/Router Channel Default
- Additional Gateway/Router Installation
- Changing Channels
- Router Depth
- Rebuild Subnet on Scanned Devices
- Scan Devices
- Chapter 2
- Introduction
- Transceiver Devices
- LED Light Indicator
- Installing Transceiver Devices
- Pull Cords/Emergency Call
- Check-in Pull Cord
- Pull Cord Transceiver with Extended Battery Pack
- Wall Mount Emergency Call
- Universal Transceiver
- Tamper
- Nurse Call
- Door/Window Transceiver
- Door/Window Transceiver with Reset Button
- PIR Sensor
- Smoke Detector
- Pendant Transceivers
- Activate the Battery
- Set up the Pendant
- Reset the Pendant
- Verify the Pendant Appears in the System
- Replace the Battery
- Test the System Operation
- Chapter 3
- Introduction
- Device Failure
- Router Failure
- Gateway Failure
- Chapter 4
- Specifications
- Power Cable Run Lengths
- Mesh Network Router/Gateway
- Quick Look Display for Quick Look Router
- Pendant Transceiver
- Pull Cord
- Nurse Call
- Door/Window Transceiver
- PIR Sensor
- Smoke Detector
- Universal Transceiver
9600 Series Wireless Call System (0510-1078-D) - Hardware Installation Guide 49
Chapter 4
Specifications
Specifications
This section contains information on maximum power cable run lengths for installation of daisy-chained
9600 Series Routers/Gateways. It also provides specifications for the Code Alert 9600 Series devices,
including details about the specific requirements and certifications.
Power Cable Run Lengths
FIGURE 4.1: Powered by CPS
FIGURE 4.2: Powered by Wall Outlet Power Supply
Number of Routers/Gateways
CPS
16/2AWG
CPS
18/2 AWG
8 390' 245'
7 450' 280'
6 525' 325'
5 630' 390'
4 785' 490'
3 1050' 650'
2 1575' 980'
1 3150' 1960'
Number of Routers/Gateways
12V @ 1.5A
18/2AWG
9.0V @ 200mA
22/2 AWG
4 600'
3 740'
2 1000'
1 2000' 125'