User Manual
Table Of Contents
- Important Warnings
- Bio-Incompatibility Notice
- Compliance
- Introduction
- About this Guide
- Additional Detailed Documentation
- Contact Information
- Product Warranty
- Chapter 1
- Introduction
- Installation Checklist
- Installing Components
- Install the Central Server
- Install the Gateway
- Install Routers
- Place a Pendant into Survey Mode
- Determine Placement of Routers
- Quick Look Router
- Gateway/Router Reset Button
- LED Sequence
- Gateway/Router Channel Default
- Additional Gateway/Router Installation
- Changing Channels
- Router Depth
- Rebuild Subnet on Scanned Devices
- Scan Devices
- Chapter 2
- Introduction
- Transceiver Devices
- LED Light Indicator
- Installing Transceiver Devices
- Pull Cords/Emergency Call
- Check-in Pull Cord
- Pull Cord Transceiver with Extended Battery Pack
- Wall Mount Emergency Call
- Universal Transceiver
- Tamper
- Nurse Call
- Door/Window Transceiver
- Door/Window Transceiver with Reset Button
- PIR Sensor
- Smoke Detector
- Pendant Transceivers
- Activate the Battery
- Set up the Pendant
- Reset the Pendant
- Verify the Pendant Appears in the System
- Replace the Battery
- Test the System Operation
- Chapter 3
- Introduction
- Device Failure
- Router Failure
- Gateway Failure
- Chapter 4
- Specifications
- Power Cable Run Lengths
- Mesh Network Router/Gateway
- Quick Look Display for Quick Look Router
- Pendant Transceiver
- Pull Cord
- Nurse Call
- Door/Window Transceiver
- PIR Sensor
- Smoke Detector
- Universal Transceiver
9600 Series Wireless Call System (0510-1078-D) - Hardware Installation Guide 35
Installing Transceiver Devices
To mount the Door/Window transceiver using mounting hole
1. Use your fingers to open the front cover of the transceiver by pushing the
retaining clip on the top downward. This allows the hinged cover to drop
down and expose the circuit board.
2. Carefully pull the circuit board free from the pins on the enclosure.
3. Drill a 1/8 inch mounting hole through the indentation in the transceiver
enclosure.
4. Hold the enclosure against the door/window frame so that when the circuit
board is in place, the internal reed switch faces the magnet on the door.
5. Mark out the mounting hole and drill a 1/16 inch diameter hole where you
made the mark.
6. Line up the hole on the transceiver enclosure with the newly drilled hole
and mount the enclosure to the door frame with a number 6 screw (not
included).
7. Pull the plastic battery tab to activate the battery or refer to the section entitled “To replace the
battery in a Universal transceiver” on page 31.
8. Return the circuit board onto the two pins.
9. Snap the front cover of the transceiver back into place.
To mount the Door/Window transceivers with adhesive pad
The Door/Window transceiver can be attached using an adhesive pad that
can be purchased with the transceiver.
1. Determine where you want to mount the transceiver. The mounting
surface must be clean, dry and free of condensed moisture.
2. Pull the plastic battery tab to activate the battery or refer to the section
entitled “Smoke Detector” on page 39 to insert a new battery.
3. Remove the adhesive plastic guard.
4. Hold the enclosure against the door/window frame so that when the
circuit board is in place, the internal reed switch faces the magnet on the
door.
5. Firmly attach the transceiver to the door/window frame. Ideal application temperature range is 70°F
to 100°F (21°C to 38°C).
6. Press firmly and hold for approximately 10 seconds to activate the pressure sensitive adhesive.
7. After application, the bond strength will increase as the adhesive flows onto the surface. At room
temperature, approximately 50% of ultimate bond strength will be achieved after 20 minutes, 90%
after 24 hours and 100% after 72 hours.
NOTE: Orientate the internal reed switch next to magnet 3/8
inch away. The Universal
transmitter has been tested and shown to work as far as 1/2 inch.
Drill hole here