9600 Series Wireless Call System (0510-1078-D) - Hardware Installation Guide 21
Chapter 2
Installing Transceiver Devices
Introduction
This chapter provides detailed information about installing transceiver devices and testing the operations of
the 9600 Series Wireless Call System with those devices. Transceivers are devices that transmit and receive
data. The transceiver devices listed in this section are supported by the Code Alert 9600 Wireless Call System
running the Series 6.0 Software application. The Code Alert 9600 Wireless Call System includes transceivers
that are carried by the patient and fixed devices. A fixed device is a stationary device that is assigned to a
room or a unit. Fixed devices are not transported with the patient but stays in the unit to which they are
assigned.
Transceiver Devices
The transceiver devices are entered into the system by placing the device into an alarming state. The system
senses the device when the device goes into alarm and adds it to the device list in the Configuration-Device
window.
The user must then update the device information; for example, give the device a name and/or enable
features, refer to the Series 6.0 System Administrator Guide (PN 0510-1080).
The Code Alert 9600 Wireless Call System transceiver devices consist of the following:
Pull Cord
Wall Mount Emergency Call
Universal
Nurse Call
Door/Window
PIR Sensor
Smoke Detector
Pendant
WA R N I N G: Before cutting openings or drilling holes through walls, you must
verify that you will not strike any wiring or plumbing.
NOTE: Pull Cords and Universal Transceivers check in every 20 minutes; Pendants
check-in every 100 seconds by default. These devices will auto populate the device list
when they check-in.