User Manual

Table Of Contents
Series 6.0 Software (0510-1065-A) - User Guide 11
9450 System
Mother Transmitter
A Mother transmitter is placed on the wrist of a new mother and then linked to an Infant
transmitter worn by her newborn. When the Infant transmitter is in proximity of the
Mother transmitter, the Mother transmitter flashes Green light to indicate that the Infant
and Mother transmitters match. If there is a mismatch, meaning the Infant and Mother
transmitters do not match, the Mother transmitter flashes the Red light three times, and
sounds a three beep alarm. A Mismatch Transmitter alarm is also generated in the
Event List at the computer. If there is a linking problem, a Link Transmitter
alarm is generated in the Event List at the computer.
In cases where the Mother Transmitter is linked but no longer active (e.g. the patient has been
discharged but the transmitter was never cleared) the transmitter will beep after 24 hours of inactivity.
If the transmitter goes undetected and allowed to beep for 15 minutes, it will clear and discharge itself
from the system. Tapping the Mother Transmitter will temporarily stop the beeping but the
transmitter remains in a linked state until manually cleared (refer to the Mother Baby Match Mother
Transmitter User Guide). If not cleared, the transmitter will beep again after another 24 hours of
inactivity.
NOTE: Although the Mother transmitter will link to an Asset transmitter, the software does
not support this functionality. If this happens, clear the mother transmitter and re-link the
infant and mother transmitters.
CAUTION: The Mother Transmitter can be worn in the shower; complete
submersion should be avoided. Prolonged submersion may result in equipment
and/or system failure and voids any or all Product Warranty.