CHAPTER 2 Installing Transceiver Devices
Quick Response Premiere Wireless Call System Hardware Installation Guide Page 37 of 62
0510-1099-C
3. Center punch each hole and install two nylon wall anchors (included). If the call station is located
on a concrete wall then you must use the wall anchors designed for use with concrete (not included).
4. Snap the front assembly over the rear plate and place a screw (included) into the bottom of the
assembly (as shown below) to secure.
Setting the Call Stations Up For Use
To set up the Nurse Call transceiver for use
1. Plug in the desired Nurse Call Cord set into the Nurse Call transceiver.
2. Activate the Nurse Call transceiver by pressing the button.
3. If the Nurse Call transceiver is working properly, an Assistance Required alarm event will be listed
on the Event List at the Central Server. The system senses the Nurse Call transceiver when it goes
into alarm and adds it to its list of devices.
Reset the Nurse Call transceiver by pulling the button back to the UP position. The alarm event will
automatically clear from the Event List at the Central Server. If Enforce Joint Commission is
enabled, you must clear the alarm at the Central Server or Client computer by selecting a reason.
4. At the Central Server, update the Nurse Call transceiver information, for example, giving the Nurse
Call a name and/or assigning it to a room or unit. Refer to the “Update Devices” section in the
applicable Series Software Administrator Guide.
To set up the Emergency Call transceiver for use
1. Activate the Emergency Call transceiver by pushing the red button. When pushed, the button
remains in indicating the device is in an alarm state.
If the Emergency Call transceiver is working properly, an Assistance Required alarm event will be
listed on the Event List at the Central Server. The system senses the Emergency Call when it goes
into alarm and adds it to its list of devices.
2. Reset the Emergency Call by pushing the red button again, the button pops out indicating the
transceiver has changes states and is now idle and ready for its next usage.
The alarm event will automatically clear from the Event List at the Central Server. If Enforce Joint
Commission is enabled, you must clear the alarm at the Central Server by selecting a reason.
3. At the Central Server, update the Emergency Call information, for example, giving the Emergency
Call a name and/or assigning it to a room or unit. Refer to the “Update Devices” section in the
applicable Series Software Administrator Guide.