User's Manual

Chapter 3 Alarms and Troubleshooting
Baby Check Transmitter User Guide Page 19 of 22
0510-1124-A
1 Amber
Baby Check Transmitter is in Link Mode
1. The button on the Baby Check transmitter is pressed once to
verify it is ready to be linked.
2. The amber light signals Link Mode.
3 Green
Baby Check Transmitter Decodes 4 Signals and Links
1. Transmitter is successfully linked.
2. No action required; operates as designed.
3 Red
Baby Check and Infant Transmitter Mismatch
1. An Infant transmitter comes in contact with the Baby Check
transmitter. The Baby Check and Infant transmitters do not
match.
2. Follow your facility’s established procedures to secure and
separate the infant from the mismatched Baby Check.
3. Clear the alarm at the Client computer (you must select an
event cause).
5 Red
Baby Check Transmitter Indicates Low Battery
1. Discard transmitter or call RFT Tech Support for replacement.
Green, Red,
Green, Red,
Green
Baby Check Transmitter Cleared
1. With the Transmitter Clearing Device 10 to 12 inches in front
of the Baby Check transmitter, the button on the Baby Check
transmitter is pushed to clear it and remove all linked activity.
2. Store the cleared Baby Check transmitter together with the
Transmitter Clearing Device.