Chapter 4: Maintenance
14 CA630 System (0510-1116-A) - In-Service Manual
To test your Nurse Call Functionality
1. Follow the set up instruction in the section Setting Up The System
With the system set up for Nurse Call, push the button on the Nurse Call cord set.
If the system is operating properly, an alert will sound at the control unit.
Operating and Storage Conditions
Pads
Pads must be stored properly to prevent damage. Store flat. Do not fold or store other items
on top of pads. Pads have a limited life; mark each pad with the WARRANTY EXPIRES
date.
Control Unit
Operating Temperature
40 °F to 100 °F (4.4 °C to 38 °C)
Operating Humidity
0-95% non-condensing
Storage Temperature
32 °F to 120 °F (0 °C to 248 °C)
Storage Humidity
0-95% non-condensing
Service and Return
Do not attempt to service or repair the CA630 System; there are no serviceable parts
inside the control unit or the sensor pad. Any attempt at servicing or repairing the
product voids the warranty.
If you encounter problems with your equipment, please contact RF Technologies for
assistance. If your equipment appears to be defective, a technician will issue a Return
Merchandise Authorization Number so the unit may be returned.