CA630 System (0510-1116-A) - In-Service Manual 8
Nurse Call
To use the Nurse Call function:
1. Connect the Nurse Call modular jack cable to the control unit.
2. Insert the Nurse Call cord set into the modular jack. Nurse Call monitoring
starts when the cord set is connected.
3. Push the button on the Nurse Call cord set, an Assistance required alarm event
is reported in the Event List at the Client computer.
4. If the cord is disconnected, an alarm will sound at the control unit and at
remote indicators.
Guidelines for Pressing the Hold Button
1. Press the Hold button to initiate a 30-second alarm pause. Press the Hold
button again to cancel the 30-second pause.
2. The system turns off if no weight is applied to the pad after the 30-second
pause. The system reactivates once weight is applied to the pad.
3. To discontinue monitoring and avoid triggering an alarm, press the Hold
button. While the control unit is in the Hold mode you can safely remove the
patient from the Sensor pad and disconnect the Sensor pad and the Nurse Call
cord.
NOTE: While in Hold mode, an Assistance Required alarm (generated by pushing
the button on the Nurse Call cord) will not alarm at the unit but will post at the
Client computer and in the Reports.