User Manual
Chapter 1: System Overview
4 CA630 System (0510-1116-A) - In-Service Manual
Control Unit
The control unit is designed to fit inside a silicone protective covering (Protective
Boot) to minimize damage from dropping and electro-static discharge. It also
includes an attachment strap for mounting the control unit to a bed or chair.
Status Indicator lights
Volume Indicator Lights
NOTE: The control unit should always be used inside the Protective Boot.
Control Unit Indicators
1. VOLUME INDICATOR LIGHTS
• As you adjust the volume, the green indicator light aluminate
to indicating the volume level.
2. STATUS INDICATOR LIGHTS
• GREEN (MONITOR ON)—The green light flashes every 2-seconds to
indicate the control unit is monitoring.
• ORANGE (HOLD)—The orange light flashes on and off to indicate the
control unit is in hold mode.
• RED (ALARMS)—The red light flashes on and off to indicate an alarm.
Alarms include Fall, Assist and Low Battery.