Troubleshooting Guidelinese
Smart Sense Infant Transmitter (0510-1093-C) - User Guide 21
Multiple No Signal
Alarms - Transmitter
IN use (applied to a
patient)
1. Ensure that the patient is safe and within the Safe Place monitored
area. If patient is not within the area - see the last section below for
Escort and Transfer, or return the patient to the monitored area.
2. Clear No Signal alarm at the Safe Place workstation.
3. Request an Adjust at the Safe Place workstation.
4. Obtain a new piece of approved banding material and return to the
bedside.
5. Replace banding material during the Adjust period and ensure that
clamps are securely fastened on both sides.
6. If a new No Signal alarm occurs for the same patient, following
steps 1 and 2, obtain a new transmitter.
7. Band the patient with the new transmitter.
8. Re-band the transmitter removed from patient with a piece of
approved banding material, ensuring that both clamps are securely
fastened.
9. Wait 60 seconds.
10. Request Discharge
11. Remove banding material
12. If Discharge expires after following steps 8-11, contact a superuser
to perform "Hard" removal of transmitter.13.
13. Admit the transmitter applied in step 7.