Chapter 1: The Smart Sense Transmitter
14 Smart Sense Infant Transmitter (0510-1093-C) - User Guide
7. Inspect the piercing contacts on the transmitter. The contacts must be gold in color and
not bent or out of alignment. If the piercing contacts are NOT gold in color or are bent
out of alignment, discard the transmitter.
8. Close the clamps on the transmitter.
9. Place the clean transmitter in a storage container for future use.
Approved Cleaners
Following is a list of a approved cleaners recommended for use when cleaning the
Smart Sense transmitter.
CAUTION: The transmitters may be sprayed or rinsed with
water or the approved cleaners during cleaning, but must not be
soaked. Soaking transmitters may result in equipment and/or
system failure.
Approved Cleaners
Airex 109A Isopropyl Alcohol Quest 256
Cavicide™ Neutral Quat 128™ Virex II-256
Health-Tec™ Quik Fill 920™ Sani-cloth HB (wipes)
Hi-Tor Plus™
NOTE: Some agents are intended for topical, antimicrobial use only and
will not destroy viruses such as Hepatitis B and HIV
Piercing Contacts
Brush