User Manual

Cleaning and Caring
Smart Sense Infant Transmitter (0510-1093-C) - User Guide 13
Cleaning and Caring
This section provides detailed information about cleaning and caring for a Smart
Sense transmitter.
To clean a Smart Sense transmitter after use
1. Remove the banding material from the transmitter and dispose of it.
2. Pre-clean the Smart Sense transmitter using an approved cleaner.
3. Use the provided brush to clean the hard-to-reach areas. Brushes must be disinfected
after each use according to the disinfectant manufacturers instructions and your
facility’s infection control procedures.
4. Disinfect the Smart Sense transmitter by spraying it with an approved cleaner.
5. Allow the cleaner to remain on the Smart Sense transmitter for the amount of time
specified by the cleaner manufacturer. When using surface disinfectants and
decontaminate cleaners, always follow the cleaner manufacturers instructions.
6. Wipe the transmitter with a clean cloth or towel, and then allow the transmitter to air-
dry. The transmitter can also be rinsed with water, if required in the cleaner
manufacturers instructions.
CAUTION: Use of non-approved cleaners on Smart Sense
transmitters can cause permanent damage to the transmitter and
the piercing contacts and will void any and all warranties.
NOTE: In order to avoid nuisance alarms, cleaning procedures must be
performed at least 100 feet from any Exit Alarm Zone.
CAUTION: Using an approved cleaner does not confirm the
suitability of the cleaner to meet your infection control standard;
it only indicates that the Smart Sense transmitter will not be
harmed by the cleaner.
CAUTION: Use care around the piercing contacts on the Smart
Sense transmitter as they are extremely sharp. Placing your
fingers directly on the piercing contact can result in injury.