The Smart Sense Transmitter
Smart Sense Infant Transmitter (0510-1093-C) - User Guide 9
2. With the yellow-side of the banding material facing outwards, use the plastic guides to
thread the banding material through the transmitter clamp slot.
3. Pull the tab until it clears the clamp and push the clamp lever down until it snaps shut.
4. Place the transmitter with banding material around the infant’s ankle and thread the
banding material through the second transmitter clamp slot.
The Smart Sense Transmitter should fit snugly against the infant’s skin without a gap
between the infant’s skin and the transmitter. Do not overtighten the band.
CAUTION: Do not thread both sides of the banding material
through the clamps until you are ready to secure the transmitter
to the infant. Pre-threading may allow the transmitter to auto-
enroll and alarm.
NOTE: You must ensure transmitter is banded properly-without an air gap
between the transmitter and skin, yet allowing the band and transmitter to be
turned on the ankle. Over-tightening the band will not reduce the number of
alarms and could cause patient injury.