Series 7.0 Software (0510-1086-B) - User Guide 33
Advanced 4-Way Care Solution
Advanced 4-Way Control Unit
The Advanced 4-Way Care control unit attaches to a patient’s bed or chair, out of reach of the patient. It
connects to a Motion sensor pad, an Incontinence Detection pad and a Nurse Call cord set. If using the control
unit for Fall Management or Re-position Reminder, connect the Motion sensor pad. If using the control unit
for Incontinence Detection, connect the Dri Tec sensor pad. To allow Nurse Call functionality, connect the
Nurse Call cord set.
The control unit is powered by a 9-volt lithium battery or by a plug-in power supply. To simultaneously
address all four applications, insert the Dual Function Adapter and plug in the Motion sensor pad and the
Dri Tec sensor pad into their respective jacks in the adapter. Then insert the Nurse Call modular jack cable
into the applicable jack on the control unit and plug the Nurse Call cord set into the modular jack.
Advanced 4-Way Care Sensor Pads
The sensor pads for the Advanced 4-Way Care Solution are placed on a patient’s bed or chair and
connected to the Advanced 4-Way Care control unit. Refer to the applicable control unit’s in-service
manual for detailed information about using the control unit with the sensor pads.
Motion Sensor Pad
The Motion sensor pad is used for Fall Management and Re-Position Reminder monitoring. They are
disposable, intended for single-patient use, and warranted for 30 days (offerings for pads are subject to
change, contact Customer Care at 1-800-669-9946 for the most current offerings). When the patient’s weight
is removed from the pad, the control unit beeps three times every second to alert caregivers. The status
indicator light flashes red in conjunction with the beeps for a visual alert.
When used for a Re-position Reminder, the Motion sensor pad connects to the control unit to allow you to
select a pre-programmed time interval to remind you to change/vary the position of a patient in their bed. At
the end of the monitoring period, the control unit beeps once every second indicating the time interval has
expired. A red status indicator light flashes in conjunction with the beeps for a visual alert.
The Re-position Reminder light on the control unit also flashes red to indicate the specific bed position that
the patient was left in at the beginning of the monitoring period.
WARNING: If the keys on the control unit are worn or damaged, you must
stop using the control unit and contact RF Technologies.
WARNING: Pad may become damaged when used on a dynamic mattress
(air, water, etc.). Correct operation should be confirmed at every patient
rotation.