User's Manual

Table Of Contents
Series 7.0 Software (0510-1086-B) - User Guide 105
Event Types
No Signal Alarm
A No Signal alarm occurs when the system is unable to locate the transmitter within the required supervision
time.
To respond to a No Signal Alarm
1. Always follow your facility’s policies and procedures to ensure patient safety and secure the area.
2. Verify transmitter is fastened correctly and without signs of damage.
For more information, refer to the Patient Transmitter User Guide (PN 0510-1013) or the Infant
Transmitter User Guide (PN 0510-1006).
3. From the Client computer, click anywhere in the No Signal Alarm Message Box to access the
Event Information window.
4. Select one of the following event causes:
Clasp open —one of the clasps on the transmitter was open.
Band cut—the banding material on the transmitter was cut.
Band worn—the banding material on the transmitter is worn and must be replaced.
Unauthorized open-—an unauthorized person opened the transmitter clasp.
Adjust—a transmitter was being adjusted without authorization.
Discharged —the transmitter was being removed in a discharge without authorization.
Te s t —the system was being tested.
Other—opens a Clear Reason window, user must enter the cause in this window.
You cannot clear an event until you select the appropriate cause for the event. The event is cleared
from the Event List of every Client computer configured to monitor the unit.
5. Multiple No Signal alarms may be an indication that Receiver coverage may be inadequate.
Contact your System Administrator.
WARNING: You must test all transmitters prior to use, and periodically
thereafter, to verify proper operation. Failure to test the transmitters
before use can result in system failure and/or an elopement or abduction.
In addition, failure to test transmitters voids the RF Technologies Product
Warranty.
NOTE: You can click Silence to stop the alarm sound at the Central Server or Client
computer. The alarm is silenced at that computer only, for the configured length of time;
however, the event still appears in the Event List. The next event automatically restarts
the alarm sound.