Product Manual

120 µL Pipette
Test Procedure
All specimens must be at room temperature before testing.
Expiration date: Check expiration date on each individual test package or outer box before using. Do not use any test past the expiration date on the label.
Verify that Sofia 2 is
set to the desired
Mode: WALK
AWAY or READ
NOW. See the
“Using Sofia 2”
section for more
information.
Dispense all of the Reagent
Solution into the Reagent
Tube. Swirl the Reagent Tube
to dissolve its contents.
Place the patient swab
sample into the Reagent
Tube. Roll the swab at
least 3 times while
pressing the head
against the bottom and
side of the Reagent
Tube. Leave the Swab in
the Reagent Tube for
1 minute.
Roll the Swab head
against the inside of the
Reagent Tube as you
remove it. Dispose of
the used Swab in your
biohazard waste.
Fill the provided Small, Clear
120 µL Fixed Volume Pipette with
patient sample from the Reagent
Tube.
To fill the Fixed Volume Pipette
with the patient sample:
a) FIRMLY squeeze the top bulb.
b) Still squeezing, place the Pipette
tip into the sample.
c) With the Pipette tip still in the
sample, release pressure on bulb
to fill the Pipette.
Proceed to the
“Using Sofia 2”
section to
complete the
test.
Swab Test Procedure (Nasal/Nasopharyngeal)
Study the Package Insert and User Manual
thoroughly before using Quick Reference
Instructions. This is not a complete
Package Insert.
IMPORTANT! Read instructions carefully before
beginning. The test procedure below is unique to the
Sofia 2 Flu + SARS Antigen FIA and may differ from other
Sofia and Sofia 2 FIA procedures.
QUICK REFERENCE
INSTRUCTIONS
For use with Sofia 2.
Rx only
1
2
3
4
5
6
7
For use under the Emergency Use Authorization
(EUA) only
For in vitro diagnostic use

Summary of content (4 pages)