When the test is complete, the results will be displayed on the Sofia screen. The results can be
automatically printed on the integrated printer if this option is selected. Test Lines will not be
visible to the naked eye.
Results: The Sofia screen will display results for the procedural control as being “valid or
invalid.” If the control is “invalid,” retest with a new patient sample and new Test Cassette.
If used on asymptomatic individuals for serial testing, a second test should be obtained with at
least 24 hours (and no more than 36 hours) between tests.
Reader Display
Interpretation
SARS: Positive
Procedural Control: Valid
Positive Test SARS
(SARS Antigen present)
SARS: Negative
Procedural Control: Valid
Negative Test for SARS
(no SARS antigen detected)
SARS:
Procedural Control: Invalid
Result Invalid
From the main menu, select
Run QC.
Follow the prompts on the screen. Scan the QC Card (located on the kit box).
Sofia prompts the user to select the desired mode (WALK AWAY or READ NOW).
Run the External Control Swabs.
Follow the Swab Test Procedure of this Quick Reference Instructions. First test the
Positive Swab, followed by the Negative Swab.
After both the Positive and Negative Swabs have been run, the results will be
displayed as “Passed” or “Failed.”
Sofia Interpretation of Results
Sofia External Quality Control (External Positive and Negative Swabs are supplied in the kit)