Product Manual

Page 32 of 41
Select to return to the Run Test screen.
Run QC
External Quality Control (QC) may be tested to demonstrate that the assay-specific
reagents, Test Cassettes and assay procedures are performing properly.
To run the External Controls, refer to the assay-specific Package Insert for each specific test
for step-by-step instructions.
QC Results
Once the QC run is complete, the system will display either a valid result, , or an invalid
result, .
Select to complete the QC process and return to the Run Test screen.
If the QC controls do not pass, repeat the external control testing process or contact the
Supervisor or Quidel Technical Support at 800.874.1517 (in the U.S.) for assistance before
testing patient specimens. See page 3 for additional contact information.
Run Calibration
Note: When running Positive and Negative controls with a qualitative test, the Positive
Control must be tested first before the Negative Control.