Tips for Batch Testing
In order to make batch testing easier, the user can prepare one or more Reagent Solution Bottles in advance of testing
samples. The user can break the ampoule inside each Reagent Solution Bottle, shake to mix the solutions, and then store
the capped Bottles on the bench top at room temperature for up to 12 hours without loss of activity before using with
Swab sample(s).
Critically important, the user should never open the foil pouch thus exposing the test Cassette to the ambient environment,
until it is ready for immediate use.
Run Test
1. Input the User ID using the handheld barcode scanner or manually enter the data using the key pad.
NOTE: If you mistakenly scan the wrong barcode, use the Arrow Buttons on the Sofia key pad to re-highlight the field.
Then simply rescan using the correct barcode, and the previous one will be overwritten with the correct barcode.
2. Input Patient ID and/or Order # using the handheld barcode scanner or manually enter the data using the key pad.
3. Press Start Test and the Sofia drawer will automatically open.
4. Verify that the correct development mode, WALK AWAY or READ NOW, has
been selected. Insert the prepared patient test Cassette into the drawer of
Sofia and close the drawer.