Product Manual

UPC-A
UPC-E
EAN – 13 (WITHOUT ISBN
TRANSLATION)
CODABAR (NOT ENCODED)
INTERLEAVED 2 of 5
IATA
TELEPEN NUMERIC
CODE 128
CODE 93
GS1 DATABAR LIMITED
GS1 DATABAR TRUNCATED
IEAN-8
CODE 39 REG
CODE 39 ITALIAN (ENCODED
WITHOUT TRANSMIT ST/SP)
INDUSTRIAL 2 of 5
S-CODE
MSI/PLESSEY (NOT ENCODED)
UK/PLESSEY (ENCODED)
EAN-128
GS1 DATABAR
OMNIDIRECTIONAL
GS1 DATABAR EXPANDED
The Opticon barcode scanner (Model OPR-2001) is shipped in the
correct configuration and is ready for use after connecting to Sofia
via the PS2 port. Additional information may be obtained at
opticonUSA.com or by contacting Opticon technical support at
1.800.636.0090 x 2127.
Contact Quidel Technical Support for assistance at 800.874.1517,
(in the U.S.). For additional contact information (page 3).
The following barcodes have been configured at time of shipment:
Appendix B System Troubleshooting
Symptom
Cause Action
Paper out error Sofia detects a
printer error or that
the printer is out of
paper
Check the printer paper
and install a new roll if
needed.
SD Card error The system detects
that the SD Card is
missing or inserted
incorrectly.
Check the SD Card slot in
the back of the unit. Install
SD Card if needed.
SD Card full The system detects
the SD Card is full.
Replace SD Card in the
back of the unit.
Barcode
Scanner does
not scan or enter
data-power
The Barcode Scanner
may be unplugged.
Check to ensure the
Barcode Scanner is
connected to the back of
the unit.
Barcode
Scanner does
not scan or enter
data-Barcode
Scanner
The Barcode Scanner
may be improperly
programmed.
Contact Quidel Technical
Support at 800.874.1517
for barcode
reprogramming or to
obtain a replacement
Barcode Scanner. For
additional contact
information (page 3).
Sofia Calibration
Check failed
Calibration Cassette
stored outside of
provided light-
protecting pouch.
Test the Calibration
Cassette. If the calibration
check does not pass or
is invalid, contact Quidel
Technical Support at
800.874.1517 to obtain a
replacement Calibration
Cassette. For additional
contact information
(page 3).
Drawer Blocked
Error
Anything on bench
top in way of drawer
opening.
Unblock drawer and press
EJECT button. If error
continues, contact Quidel
Technical Support at
800.874.1517. For
additional contact
information (page 3).
Barcode Scanner
Barcode Scanner Appendix B System Troubleshooting
54
55