Product Manual
Surveillance Mode
Note: Before setting up Sofia to Sofia Surveillance Mode, please set
up other features on Sofia, particularly the Location Name for
surveillance purposes.
Sofia and Virena Router Connection Installation
Please refer to the Sofia Surveillance Mode Installation Instructions
that are supplied in the separate Viena Router package.
Set Network Configuration
Refer to the Network Configuration section (page 41).
Scroll to Set IP Address and ensure Manual is checked, see Figure 20.
Scroll to Network Address and enter IP network address to 192.168.0.5.
Verify that all configurations on Sofia including: Subnet Mask, Gateway
and Port Number match what is shown in Figure 20. If it does not match,
please enter the correct numbers. Then select OK.
Wait 1 minute, then go to View Network Settings and ensure that you
have a valid IP address.
Figure 20
Surveillance Mode Set Up
Note: If Sofia does not show Surveillance Settings as an option, Sofia
will require an upgrade to firmware 1.3.1 or above.
Under Set SURV Parameters, select the following using the Left/Right/
Up/Down arrows:
Age Required:
✔
Yes
Transmit Patient ID:
✔
Off
Auto Send:
✔
On
Queue Results:
✔
On
When completed, screen will display as Figure 21.
Figure 21
Press OK to return to the Settings screen.
Set Default
Mode &
Print Options
Set
Calibration
& QC
Set User
Access Rights
General
Settings
Surveillance
Settings
Set Entry
Requirement
& Timeout
Supervisor
Menu
Main Menu
Settings
Installation and Setup
Installation and Setup44
45