Product Manual

Network Configuration
For the Network Configuration, use the Up/Down arrows to select the
field to modify.
For the IP (internet provider) Address, use Left/Right arrows to select
β€œManual” or β€œDHCP.” Use the numeric keypad to enter address and
numbers for the Network Address, Subnet Mask, Gateway and Port
Number (Figure 18). Select OK to save changes and exit.
Figure 18
Figure 17
Continue adding the appropriate extracted specimen to its appropriate
cassette 1 minute after each until all cassettes have been charged with
specimen.
When the timer goes off for cassette #1, if required, input or scan the
User ID, patient ID and order # into Sofia for cassette #1. Then insert
cassette #1 into Sofia, Sofia will display the result in approximately
1 minute. The tray will then open. Sofia screen will display the identifier
information and results (Figure 17).
Remove cassette #1. Press Start New Test and input the required
identifier information, e.g., User ID, Patient ID and order # for cassette #2.
Place cassette #2 into the tray and then gently close the drawer. Sofia
will now automatically run test when the drawer is closed. Proceed in
the same fashion until all cassettes have been tested.
Supervisor
Menu
Main Menu
Set Location
Name
Set Data Set Time
Set Language
& Sound
Set LIS
Parameters
Settings
General
Settings
Set Network
Configuration
Installation and Setup
Installation and Setup40
41