SofiaInfluenzaA+BFIA
Page9of25
NOTE:Toobtainaccurateresults,avoidmucoidsubstanceswhen
fillingtheLarge,PinkFixedVolumePipettewithpatientsample
fromthecollectioncup.
4. FirmlysqueezethetopbulbtoemptythecontentsoftheLarge,PinkFixed
VolumePipetteintotheReagentTube.Extraliquidleftoverintheoverflow
bulbshouldbeleftbehind.
NOTE:OncethesampleisaddedtotheReagentTube,vigorouslymixpriorto
addingthesampletothetestCassette.
NOTE:TheFixedVolumePipettesaredesignedtocollectanddispensethe
correctamountofliquidsample.DiscardthePipetteinyourbiohazardwaste.
5. FilltheprovidedSmall,Clear120µLFixedVolumePipettewithpatientsample
fromtheReagentTube,byslowlyreleasingpressureonthebulb.
6. Firmlysqueezeth
etopbulbtoemptythecontentsoftheSm all,ClearFixed
VolumePipetteintotheCassettesamplewell.Extraliquidleftoverintheoverflow
bulbshouldbeleftbehind.DiscardthePipetteinyourbiohazardwaste.
7. Promptlypro
ceedtothenextsection,“UsingSofia,”tocomp letethetest.
USINGSOFIA
Toobtainaccurateresults,SofiaandthetestCassetteshouldnotbeleftstandinginsidealaminarflowhood
orinaheavilyventilatedareaduringthe15‐minutedevel opmentstep.Theaccentuatedairflowcandrythe
sample and retard the flow of the sample through the test strip which can cause inaccurate or in
valid
results.
Ifinfectionbyanovelvirusissuspectedorifyourinstitution’spolicyrequiresperformingrapidinfluenzatests
insidealaminarflowhood,afteraddingthesampletothetestCassette,coverthetestCassetteorinsertit
backintothefoilpouchuntilitisreadytoinsertin
toSofia.