Product Manual

_________________________________________________________________________________
SofiaInfluenzaA+BFIA
 Page4of25
QUALITYCONTROL
TherearethreetypesofQualityControlforSofiaandtheCassette:SofiaCalibrationCheckprocedure,built
inproceduralcontrolfeatures,andExternalControls.
SofiaCalibrationCheckProcedure
Note:Thisisa“CalibrationCheck”procedure.
TheCalibrationCheckProcedureshouldbeperformedevery30days.Sofiacanbeeasilysettoremindtheuser
tocompletetheCalibrationCheckProcedure.
TheCalibrationCheck is a requiredfunctionthatchecks Sofiaoptics andcalculationsystemsusingaspecific
Calibration Cassette. Th
is Calibration Cassette is supplied with the Sofia installation pack. Refer to the Sofia
UserManualfordetailsregardingtheC alibrationCheck Procedure.
Important: Ensure that the Calibration Cassette is stored in the provided storage pouch between uses to
protectfromexposuretolight.
1. To check the calibration of Sofia, select “Calibratio
n” from the
MainMenu.
2. Followingtheprompts,inserttheCalibrationCassetteintoSofia
and close the drawer. Sofia performs the Calibration Check
automaticallywithintwominuteswithnouserinputrequired.
Sofia indicates when the Calibration Check is completed. Select OK to
returnto
theMainMe
nu.
NOTE:IftheCalibrationCheckdoesnotpass,notifytheonsiteSupervisororcontactQuidelTechnicalS upport
for assistance Monday through Friday from 7:00 a.m. to 5:00 p.m. Pacific Time at 800.874.1517 (in
the U.S.); 858.552.1100 (outside the U.S.); Fax: 858.455.4960; custserv@quidel.com (Customer Service);
technicalsupport@quidel. com(TechnicalSup
port);orcontactyourlocaldistributor.