Product Manual
_________________________________________________________________________________
SofiaInfluenzaA+BFIA
Page4of25
QUALITYCONTROL
TherearethreetypesofQualityControlforSofiaandtheCassette:SofiaCalibrationCheckprocedure,built‐
inproceduralcontrolfeatures,andExternalControls.
SofiaCalibrationCheckProcedure
Note:Thisisa“CalibrationCheck”procedure.
TheCalibrationCheckProcedureshouldbeperformedevery30days.Sofiacanbeeasilysettoremindtheuser
tocompletetheCalibrationCheckProcedure.
TheCalibrationCheck is a requiredfunctionthatchecks Sofiaoptics andcalculationsystemsusingaspecific
Calibration Cassette. Th
is Calibration Cassette is supplied with the Sofia installation pack. Refer to the Sofia
UserManualfordetailsregardingtheC alibrationCheck Procedure.
Important: Ensure that the Calibration Cassette is stored in the provided storage pouch between uses to
protectfromexposuretolight.
1. To check the calibration of Sofia, select “Calibratio
n” from the
MainMenu.
2. Followingtheprompts,inserttheCalibrationCassetteintoSofia
and close the drawer. Sofia performs the Calibration Check
automaticallywithintwominuteswithnouserinputrequired.
Sofia indicates when the Calibration Check is completed. Select OK to
returnto
theMainMe
nu.
NOTE:IftheCalibrationCheckdoesnotpass,notifytheon‐siteSupervisororcontactQuidelTechnicalS upport
for assistance Monday through Friday from 7:00 a.m. to 5:00 p.m. Pacific Time at 800.874.1517 (in
the U.S.); 858.552.1100 (outside the U.S.); Fax: 858.455.4960; custserv@quidel.com (Customer Service);
technicalsupport@quidel. com(TechnicalSup
port);orcontactyourlocaldistributor.