Product Manual

Page 8 of 124QuickVue RSV Test
FOR INFORMATIONAL USE ONLY
FOR INFORMATIONAL USE ONLY
Not to be used for performing assay. Refer to most current package insert accompanying your test kit.
Nasopharyngeal Aspirate or Nasal/Nasopharyngeal Wash Test
Procedure
1. Just before testing, add Extraction Reagent to the test tube up to
the fill line (250 µL).
Note: Too little or too much of the Extraction Reagent may cause
erroneous results
2. To fill the pipette with the sample*:
a) FIRMLY squeeze the top bulb.
b) Still squeezing, place the pipette tip into the liquid sample.
c) With the pipette tip still in the liquid sample, release pressure on
bulb to fill the pipette (extra liquid in the overflow bulb is OK).
*NOTE: The pipette is designed to collect and dispense the
correct amount of liquid sample.
3. To add the sample to the test tube:
a) Firmly squeeze the top bulb to add the sample in the
pipette to the test tube with the reagent. The correct
amount will be added, even though the overflow bulb
will not empty. Discard the pipette.
b) Swirl or shake the tube to mix.
c) Wait one (1) to two (2) minutes to allow the mixture
to react.
4. Place the Test Strip into the tube with the
arrows pointing down. Do not handle
or remove the Test Strip for fifteen (15)
minutes.
5. Remove the Test Strip, and read the result according to the Interpretation
of Results section. Some positive results may appear sooner than 15 minutes.
Fill to Line
Squeeze
here
pipette
patient
sample
overflow
Squeeze
here
Swirl or
shake
Wait
1
a) b) c)