Table 2: In this study, a total of 511 serum, plasma and whole blood samples were tested
using QuickVue+ Infectious Mononucleosis test and a commercially available slide
hemagglutination test.
Of the 511 samples, 75 were found to be positive by the slide hemagglutination test
and 74 were also positive by the QuickVue+ Infectious Mononucleosis test; similarly,
436 were found to be negative by the slide hemagglutination test and 435 were also
negative by the QuickVue+ Infectious Mononucleosis test.
Based on this data, specificity was 99.8% (435/436), and sensitivity was 98.7% (74/75).
Overall agreement was 99.6% (509/511).