Quick Start Manual
Page 3
SuperLoader Bar Code Reader Upgrade Quick Start Guide
c. Carefully extend the strain relief mechanism on the bar code
reader zero insertion force (ZIF) connector.
d. Insert the end of the flex cable into the ZIF connector as far
as it can go. Then push in on the strain relief mechanism to
secure the connection.
Strain relief
(extended)
ZIF connector
ZIF connector
(with strain relief
open)
Flex cable
ZIF connector (with
strain relief closed)
e. Slide the bar code reader into the bay and do the following:
• Make sure the washer and attached screw fit into the slot
on the wall of the bar code reader bay.
• Make sure the bar code reader is seated in the pocket of
the bay. The top of the alignment tab on the bar code
reader should be flush with the top of the bar code
reader bay wall.
• Tighten the attached screw while maintaining downward
pressure on the bar code reader. Tighten just enough
to secure the bar code reader in the bay; do not
overtighten.
Washer and
attached
screw
Alignment tab
Bar code
reader bay
Bar code reader
(in correct
orientation)
Alignment tab
Washer and
attached screw
Alignment tab
(flush with
wall)
Washer and screw
(positioned in slot)
Bar code reader
(installed)
Slot
Caution: Before replacing the cover, place a straight
edge across the DCA fan cover and the top of the
bar code reader. The bar code reader must be flush,
or slightly below the level of the DCA fan shroud to
avoid damage to the bar code reader.
Verify
level with a
straight edge
across DCA
and bar code
reader