User's Guide
66
RDX
8000
Slot UI Test
Figure 41 — Slot UI Test
To execute the “Slot UI Test” select the “Start Test” button. It which will request the operator to push each
individual
slot eject button and observe the
LED
behavior. The
LED
will blink three times green then three times amber. Note
that
when the
RDX
8000 is in the slot UI test mode any inserted cartridge will not be ejected by the
associated
button
push during the test. Once all eight positions have been verified confirm completion of a
successful
test by selecting
“Test
Passed”
button which will create a “Slot UI
Test
passed” status.
Display Test
Figure 42 — Display Test
To execute the “Display Test” select the “Start Test” button. The local operator will then be asked to push the OCP
navigation button and observe the display.
Each
navigation button push will step the unit through a different display
action. The first four steps will change the display backlight from white, to red, to green, to blue verifying proper
oper
a
tion of the
RGB
backlight. The next two steps will first clear all display pixels (the observer should
see
no dots)
then set all pixels black (the observer should
see
no missing black dots in the
field).