User's Manual

Technical Specifications
DLT2000/DLT2500/DLT2700 Cartridge Tape Subsystem A-21
A.6.5 Radiated Susceptibility
Table A-23 lists the radiated susceptibility limits:
Table A-23 High Frequency, Electric Fields, 1 to 1000 Mhz
3 V/m (rms) 80% modulated 1 kHz No errors, no screen distortion
S/W recoverable errors
No hardware failure
A.6.6 Conducted Susceptibility
The transient voltage is the actual peak voltage above the normal ac voltage from the
power source.
Table A-24 lists the voltage limits for power and data cables:
Table A-24 Fast Transient (Bursts) for Power and Data Cables
2 kV S/W recoverable errors
No hardware failures
Table A-25 lists power cable voltage limits:
Table A-25 High Energy Transient Voltage for Power Cables
1.2 kV No errors
2.5 kV S/W recoverable errors
No hardware failure
_____________________________ Note_________________________
Maximum energy in a single pulse from the transient generator must be
limited to 2.5 W.
___________________________________________________________
Table A-26 lists the low-level conducted interference voltage limits:
Table A-26 Low-level Conducted Interference
3 V(rms) 80% modulated 1 kHz No errors
S/W recoverable errors
No hardware failure