22 Cabling the Library
Terminator Power 0
External terminators must be installed at the end of each SCSI bus. 
Power for the terminator must be supplied by at least one device on 
the SCSI bus.
In the factory default configuration, the library robotics supplies 
terminator power. You can turn this terminator power off using the 
Terminator Power option on the GUI Configuration screen.
Power Cabling 0
To connect the library to AC power:
1 Referring to table 2, select appropriate AC distribution unit(s) for 
your library configuration and input voltage
.
Table 2 AC Distribution 
Unit Requirements, 200V 
to 240V
Warning: AC distribution units (power strips) must be certified 
for safety by an agency acceptable to the country of 
use. For use in the United States and Canada, 
distribution units must be UL listed and CSA 
certified. For use in Europe, distribution units must 
be CE marked. National certifications for other 
countries depend on local requirements.
# of 
Library 
Modules
Distribution Unit Requirements
200-240V 50-60Hz supply voltage (Europe 
& international, etc.)
100-120V 50-60Hz supply voltage (USA, 
Canada, Japan, etc.)
1-5 Connect libraries to one AC distribution unit 
rated minimum 10A, minimum 1.00mm
2
<HAR> cord with minimum 10A plug. Must 
be 3 wire grounding plug and wall outlet.
Connect libraries to one UL Listed AC 
distribution unit rated 15A or 20A, with a 
minimum 14AWG 3 wire cord, and a 15A or 
20A plug. Must be 3 wire grounding plug and 
wall outlet.