- Quantum Cascade Laser Starter Kit Instructions Manual

Starter Kit Instruction Manual Use 33
[5] Turn the TPG128 ON (8) and set the pulse
period to 2.0 and the range to 0.5 to 10.5 µs
(medium), corresponding to a period length
of about 2.5us, or to the values specified on
the datasheet or by Alpes Lasers.
[6] Set the pulse dura tion (11) to 3.0, corre-
sp onding to a pulse length of about 50ns, or
to the values specified on the datasheet or by
Alpes Lasers.
[7] Verify laser cabling, power meter range
set corresponding to the powe r range of the
laser (specified in the datas hee t), ze roed and
aligned.
[8] Turn the HV power supply O N, then the
TPG128 ON, and slowly incr e ase the HV out-
put voltage (but not higher than specified in
the datasheet of the laser!)
[9] If available, monitor the averaged output
to avoid overloading the laser. In any case
place one probe on b oth the ano de and the
cathode of the laser and view it with an oscil-
loscope.
[10] Subtract the two traces and you get the
voltage across the las e r that should not be
more than the specified max voltage.
Note: The laboratory housing LLH100 makes
available two outputs giving access to these
voltages (see documentation).
[11] Set the desired pulse and period length as
sp e c ified in the datasheet or by special agree-
ment with Alpes Lasers.
T
P
G
1
2
8
T
T
L
P
u
l
se
G
e
n
e
r
a
t
o
r
O
F
F
O
N
P
o
w
e
r
0
.
2
t
o
2
.
2
m
s
0
.
5
t
o
1
0
.
5
m
s
5
t
o
1
0
5
m
s
0
t
o
2
0
0
n
s
P
e
r
i
o
d
D
u
r
a
t
i
o
G
a
t
e
I
N
T
r
i
g
O
U
T
O
u
t
1
O
u
t
2
5
0
o
h
m
s
5
0
o
h
m
s
5
0
o
h
m
s
m
a
x
6
0
m
A
+
1
2
V
0
V
8
9
10
11
Fig.23: Starting the TPG128