User`s manual
K e yw ord P a ra m e te r
Std /
Ne w
C om m e nts
:S YS Tem Std
:S TATe? Ne w
Q ue ry o nly. Re turns the sta te of the m achine :
re turns "1 " if the m achine is arm e d and /or
ge ne ra ting pulse s o r "0 " if the m a chine ha s be en
disa rm ed .
:B E E P e r Std S ub syste m . C o ntro ls the audib le be e p e r.
:S TATe <b oo le a n va lue > Std E na ble s/d is a ble s the be ep er.
:V O L um e <num e ric va lue > Std
S e ts the vo lum e of the bee pe r. Ra ng e is 0 to 10 0,
whe re 0 is off a nd 10 0 is m axim um volum e .
:C O M M unica te Std
S ub syste m . C o ntro ls the RS 23 2 and G P IB
inte rfa ce s.
:G P IB Std
S ub syste m . C o ntro ls the physica l co nfigura tion o f
the GP IB p ort.
:A D D Re ss
<num e ric va lue >
Std S e ts the G P IB o f the instrum ent.
:S E Ria l Std
S ub syste m . C o ntro ls the physica l co nfigura tion o f
the RS 2 32 po rt.
:B A UD
48 00 /
96 00 /
19 20 0 /
38 40 0 /
57 60 0 /
11 5 2 0 0 /
Std
S e ts the ba ud rate fo r b oth rec e iving and
tra nsm itting using the D B 9 RS 23 2 p o rt.
:US B
48 00 /
96 00 /
19 20 0 /
38 40 0 /
57 60 0 /
11 5 2 0 0 /
Ne w
S e ts the ba ud rate fo r co m m unica tion w he n using
m a pp ed co m po rts fo r U S B and E therne t
co m m unica tio n. M ust b e se t to the de fault valve
(11 5 2 0 0 ) fo r ra w TC P /IP co m m unica tio n (i.e .
La bvie w ).
:E C H o <b oo le a n va lue > N e w
E na b le s/D isa bles tra nsm issio n o f cha ra cte rs
re ce ive d o n the D B 9 se ria l p ort.
:K LO C k <b oo le a n va lue > Ne w Lo ck s the ke ypa d.
:A UTorun <b o o le an va lue > N ew
A fter p o we r-up , unit w ill sta rt g ene ra ting puls e s
auto m atica lly.
:V E RS io n? Std
Q ue ry o nly. Re turns S C P I versio n num be r in the
form : YYYY.V e x. 1 9 9 9 .0
:SERN ? N ew Q uery only. Returns the serial number.
9520 SCPI Command Summary