User guide

Table Of Contents
CAUTION! PHYSICAL INJURY HAZARD. Do not remove the covers to the
QuantStudio
®
3D Instrument or thermal cycler. There are no components inside
the instruments that you can safely service yourself. If you suspect a problem,
contact technical support (see “Customer and technical support“ on page 121).
CAUTION! PHYSICAL INJURY HAZARD. During instrument operation, the
sample block can be heated to 100°C. Before performing the following
procedure, be sure to wait until the sample block reaches room temperature.
CAUTION! Before using a recommended decontamination method other than
that listed here, contact technical support to confirm that the proposed method
will not damage the equipment.
IMPORTANT! Wear powder-free gloves when you perform this procedure.
1.
Power off and unplug the QuantStudio
®
3D Instrument or thermal cycler.
2.
Open the QuantStudio
®
3D Instrument chip tray or open the thermal cycler
heated cover to expose the sample block.
3.
If you are cleaning the thermal cycler, remove the Chip Adapters and
QuantStudio
®
Thermal Pad from the sample block.
4.
If the contamination remains, clean the chip tray or sample block using
isopropanol:
a.
Pipet a small volume of 10% isopropanol solution onto surface that contacts
the consumable, then use a cotton swab to scrub the surface.
b.
Pipet a small volume of deionized water onto surface, then use a cotton
swab to scrub the surface.
c.
Use a lint-free cloth to absorb the excess deionized water.
IMPORTANT! Always use deionized water to rinse after cleaning with
isopropanol solution.
5.
Close the heated cover or chip tray, then plug in and power on the QuantStudio
®
3D Instrument or thermal cycler.
Clean the chip tray
or sample block
Appendix A Maintenance
Cleaning the chip tray and sample block
A
QuantStudio
®
3D Digital PCR System User Guide
87