User guide
Table Of Contents
- QuantStudio 3D Digital PCR System User Guide (Pub. no. MAN0007720 Rev. C.0)
- Copyright Page
- Contents
- About this guide
- Chapter 1. Getting Started
- About the QuantStudio® 3D Digital PCR System
- Operating the instrument
- Prepare the QuantStudio® 3D Digital PCR System for use
- Connect the QuantStudio® 3D Instrument to a network
- Connect the QuantStudio® 3D AnalysisSuite™ Software
- Chapter 2. Prepare Samples and Load Reactions
- Chapter 3. Perform the PCR
- Chapter 4. Analyze the Prepared Chips
- Chapter 5. Troubleshooting
- Appendix A. Maintenance
- Appendix B. Networking
- Appendix C. Parts and Materials
- Appendix D. Specifications and Layout
- Appendix E. Safety
- Documentation and support
- Back Cover

If a data transfer is unsuccessful, you can use the QuantStudio
®
3D Instrument Run
History screen to transfer experiment files to a USB drive for manual upload to the
QuantStudio
®
3D AnalysisSuite
™
Software. Under normal conditions, the instrument
automatically transfers the results for each imaged chip immediately following the
analysis (to the cloud storage service, network file server, or USB drive); however, if
the network connection fails or if the USB drive is full, you can transfer the files
manually as follows to make them available for secondary analysis.
1.
Insert a USB drive into the port on the front of the QuantStudio
®
3D Instrument.
2.
In the touchscreen, touch
in the Main Menu.
3.
In the Settings Menu, touch Run History to view the results for the past
experiments.
Note: The QuantStudio
®
3D Instrument can store approximately 600 experiment
files before it begins purging the files from memory. When the instrument
reaches the limit, it replaces the oldest file exported from the cache with the data
from the current run.
4.
In the Run History screen, touch the entry of the experiment that you want to
transfer or view.
Transfer the
results
Chapter 4 Analyze the Prepared Chips
Analyzing the data
4
76
QuantStudio
®
3D Digital PCR System User Guide