The following materials are required to image loaded Digital PCR 20K Chips on the
QuantStudio
®
3D Digital PCR System.
Prepared Digital PCR 20K Chip (from the previous section)
Isopropanol
Gloves, powder-free, nitrile
Cleanroom-grade, low-lint polyester wipes
For optimal efficiency when running large numbers (>6) of Digital PCR 20K Chips,
follow the guidelines below:
Run the Digital PCR 20K Chips in alphanumeric order (according to chip ID) to
help avoid mistakes when entering sample information.
If you are running multiple chips, you can run the next chip in the series when
the touchscreen displays the Analyzing Chip screen .
To prepare the instrument for imaging data, edit data destination and file name
settings before loading a chip into the instrument.
1.
Set the destination for the imaging data:
a.
From the Main Menu of the QuantStudio
®
3D Instrument, touch , then
touch Data Destinations.
b.
In the Data Destinations screen, touch the desired destination for the
imaging data.
You can configure the QuantStudio
®
3D Instrument to upload the imaging
data to a USB drive, a network file server, or the cloud data storage service.
Materials
required
Guidelines for
high-throughput
analysis
Prepare the
instrument for
imaging data
Chapter 4 Analyze the Prepared Chips
Analyze the chips
4
QuantStudio
®
3D Digital PCR System User Guide