User guide

Table Of Contents
4.
Use the thermal cycler to select and start the pre-programmed run for the Digital
PCR 20K Chips. See the ProFlex
PCR System User Guide (Pub. no. MAN0007697)
or the GeneAmp
®
PCR System 9700 Base Module User Manual (Pub. no. 4303481) for
more information on running methods.
You can remove the Digital PCR 20K Chips from the block immediately after the
final extension step at 60°C is complete and the temperature of the block is
<25°C.
IMPORTANT! Read the Digital PCR 20K Chips within one hour after you
remove them from the sample block. After thermal cycling, the Digital PCR 20K
Chips can remain on the sample block for up to 24 hours at 10°C.
Note: If you programmed your thermal cycler to perform a 10°C hold following
thermal cycling, do the following to prevent condensation:
1.
Confirm the final extension step at 60°C is complete, then stop the run:
·
ProFlex
System – Press Stop Run, then press OK.
·
GeneAmp
®
PCR System 9700 – Press STOP twice , then press Exit.
2. Allow the thermal cycler to sit for at least 10 minutes with the heated lid
closed.
3. Remove the chips.
Use a low-lint wipe sprayed with isopropanol to remove any remaining
condensation or immersion fluid from the lids before imaging them.
CAUTION! PHYSICAL INJURY HAZARD. During operation, the sample
block can be heated to 100°C. Before performing the following procedure, be
sure to wait until the sample block reaches room temperature.
1.
Open the heated cover to expose the Digital PCR 20K Chip
s.
2.
Remove the QuantStudio
®
Thermal Pads from the sample block and set them on
a clean, dry surface.
IMPORTANT! Do not discard the
QuantStudio
®
Thermal Pads. They can be
reused multiple times.
Unload the
thermal cycler
Chapter 3 Perform the PCR
Thermal cycle the Digital PCR 20K Chips
3
64
QuantStudio
®
3D Digital PCR System User Guide