3.
Load the QuantStudio
®
3D Digital PCR 20K Chips into the thermal cycler:
a.
Place the sealed Digital PCR 20K Chips onto the sample block so that the fill
ports on the chips are positioned toward the front of the thermal cycler. The
fill port must be elevated during thermal cycling to prevent leaks from
improperly sealed chips.
IMPORTANT! If you are thermal cycling less than 24 Digital PCR 20K
Chips, load according to the following guidelines:
·
Load each Digital PCR 20K Chip so that the fill port is oriented toward
the front of the thermal cycler.
X0300036
Front of instrument
CAUTION HOT
24x
·
Load the right sample block first, placing at least 1 chip on the right
sample block.
·
Balance the load between the left and right blocks so that the pressure
applied by the heated cover and thermal pads is uniform across all of the
loaded chips.
b.
Lay the QuantStudio
®
Thermal Pad
s over the Digital PCR 20K Chips.
IMPORTANT! Make certain that each QuantStudio
®
Thermal Pad is
centered on the sample block and that it completely covers the Digital PCR
20K Chips.
IMPORTANT! Do not allow the QuantStudio
®
Thermal Pads to become
dirty. Fluids, debris, and other contaminants that adhere to the pad can
transfer to the
Digital PCR 20K Chips during thermal cycling and fluoresce
during imaging.
CAUTION HOT
2x
c.
Close and engage the heated cover with the QuantStudio
®
3D Digital PCR
20K Chip
s.
IMPORTANT! Verify the heater cover is completely engaged to ensure a
proper seal.
Chapter 3 Perform the PCR
Thermal cycle the Digital PCR 20K Chips
3
QuantStudio
®
3D Digital PCR System User Guide