Seal the Digital PCR 20K Chip:
a.
Remove the red protective film from the QuantStudio
®
3D Digital PCR Chip
Case Lid, then carefully place it onto the prepared Digital PCR 20K Chip so
that the window in the lid aligns with the chip.
IMPORTANT! Before applying the Chip Case Lid, make certain that it is
correctly oriented to the Digital PCR 20K Chip assembly.
b.
Pinch together all four sides and corners of the
Chip Case for at least
10 seconds to seal it. While pressing, apply some lateral pressure to ensure a
tight seal.
Note: You can remove the Digital PCR 20K Chip from the loading surface
when you apply pressure to seal it.
X0300036
X0300036
>10 sec.
8.
Fill the Chip Case with Immersion Fluid:
a.
Hold the Chip Case by its edges and at a 45° angle so that air can escape
from the loading port as you fill it, then carefully dispense Immersion Fluid
into the port until the
Chip Case contains an air bubble no larger than the fill
port (<2-3 mm in diameter).
After filling the Immersion Fluid, rotate it slightly to reveal any air bubbles
that might be hidden behind the serial number label. If bubbles are present,
add additional Immersion Fluid to remove them.
IMPORTANT! Expel the Immersion Fluid slowly. If injected too quickly, the
fluid can flush the reactions suspended within the chip.
Note: Minimize air bubble creation when you dispense the fluid; however,
one small air bubble (<2-3 mm in diameter) is acceptable.
X0300036
X0300036
b.
Using a low-lint wipe, remove any excess
Immersion Fluid from the Chip
Case to ensure optimal imaging on the QuantStudio
®
3D Instrument.