c.
While holding the Sample Loading Blade at a 70-80° angle relative to the
heated block so that the port faces up, place the edge of the blade at the end
of the Digital PCR 20K Chip. Adjust the angle of the blade until you visually
confirm that it is wetting the chip. Then, in one smooth motion, slowly drag
the blade across the chip while applying a slight downward pressure to
dispense the reaction.
IMPORTANT! Move the Sample Loading Blade slowly. The time required
to move the blade across the chip should be approximately 10-12 seconds.
IMPORTANT! Before you draw the blade across the chip, visually confirm
that the
Sample Loading Blade is wetting the chip. If not, adjust the angle
slightly.
40±1°C
70-80°
10 sec.
d.
Confirm that the Sample Loading Blade contains no residual fluid,
indicating that the reaction has been completely loaded onto the chip.
IMPORTANT! Do not apply the Sample Loading Blade
to the Digital PCR
20K Chip more than once.
e.
Allow the Digital PCR 20K Chip to sit on the heated block for 20 seconds
while the PCR reaction stabilizes within the chip.
f.
Add several drops of
Immersion Fluid directly to the chip so that the fluid
covers the entire surface.
IMPORTANT! Do not allow the syringe tip to contact the
Digital PCR 20K
Chip.
Correct
Fluid on
edges
Incorrect
Insufficient
fluid
Fluid over
chip
Note: A small amount of fluid running off the surface of the chip is
acceptable.
Chapter 2 Prepare Samples and Load Reactions
Load chips manually
2
54
QuantStudio
®
3D Digital PCR System User Guide