Remove the following consumables from their packaging and place them on a
clean, dry, lint-free surface:
QuantStudio
®
3D Digital PCR Chip Case Lid
QuantStudio
®
3D Digital PCR Sample Loading Blade
5.
Open the QuantStudio
®
3D Digital PCR 20K Chip package, gently grasp the chip
by its sides, then place it face-up on the heated block, preheated to 40°C.
IMPORTANT! Allow the
Digital PCR 20K Chip to sit on the heated block for at
least 5 seconds before loading it.
40±1°C
6.
Load the PCR reaction onto the Digital PCR 20K Chip:
a.
Briefly vortex and centrifuge the prepared reaction (master mix, assay, and
template), then aspirate 14.5 μL of the solution into a P20 pipette. If the
reactions were placed on ice, allow them to warm prior to loading.
b.
While holding the Sample Loading Blade
by its edges, carefully dispense the
reaction into the sample-loading port. After loading, gently tap the top of
the Sample Loading Blade to spread the reaction across the full length of the
blade.
While filling the Sample Loading Blade:
Hold the pipette so that the tip meets the sample-loading port at a
sharp angle and so it does not deflect either blade (forward or rear).
Do not depress the pipette to the second stop. Doing so increases the
chance of introducing an air bubble to the Sample Loading Blade.
Avoid creating air bubbles.
Chapter 2 Prepare Samples and Load Reactions
Load chips manually
2
QuantStudio
®
3D Digital PCR System User Guide
53