User guide
Table Of Contents
- QuantStudio 3D Digital PCR System User Guide (Pub. no. MAN0007720 Rev. C.0)
- Copyright Page
- Contents
- About this guide
- Chapter 1. Getting Started
- About the QuantStudio® 3D Digital PCR System
- Operating the instrument
- Prepare the QuantStudio® 3D Digital PCR System for use
- Connect the QuantStudio® 3D Instrument to a network
- Connect the QuantStudio® 3D AnalysisSuite™ Software
- Chapter 2. Prepare Samples and Load Reactions
- Chapter 3. Perform the PCR
- Chapter 4. Analyze the Prepared Chips
- Chapter 5. Troubleshooting
- Appendix A. Maintenance
- Appendix B. Networking
- Appendix C. Parts and Materials
- Appendix D. Specifications and Layout
- Appendix E. Safety
- Documentation and support
- Back Cover

• Immersion Fluid Tip
•
UV-Activated Chip Sealant Syringe
• UV-Curing Stylus for Chip Sealant
• UV-Curing Stylus Stand
• Isopropanol
• Gloves, powder-free, nitrile
• Heated block (capable of maintaining 40±1°C)
• Scissors
• Cleanroom-grade, low-lint polyester wipes
• Microcentrifuge
• Pipettes and tips, P10 to P1000
• Vortexer
WARNING! ULTRAVIOLET LIGHT HAZARD. Looking directly at a UV light
source can cause serious eye damage. Never look directly at a UV light source
and always prevent others from UV exposure. Follow the manufacturer’s
recommendations for appropriate protective eyewear and clothing.
IMPORTANT! Wear powder-free gloves while preparing the QuantStudio
®
3D
Digital PCR 20K Chips.
1.
Prepare the workspace:
a.
Preheat the heated block to 40°C, the optimal temperature for loading
Digital PCR 20K Chips.
Note: The loading surface should be a flat, temperature-controlled surface
that is large enough to accommodate the Digital PCR 20K Chip consumable.
b.
Allow the prepared digital PCR reaction to equilibrate to room temperature
(approximately 15 minutes).
c.
If necessary, replace the batteries of the UV Stylus in the
UV Stylus Stand.
Note: Replace the batteries of the UV Stylus after 10 hours of use.
2.
Prepare the syringe containing Chip Sealant:
a.
Remove the syringe, plunger, and tip from the protective packaging.
IMPORTANT! Do not discard the brown plastic bag provided. When not in
use, store the Chip Sealant syringe within the bag and in a dark location.
b.
Remove the protective caps from both ends of the syringe, push and twist
the tip to lock it into place, then insert the plunger into the opposite end of
the syringe.
c.
Place the assembled syringe within its package until ready for use.
Load and seal the
Digital PCR 20K
Chips
Chapter 2 Prepare Samples and Load Reactions
Load chips manually
2
QuantStudio
®
3D Digital PCR System User Guide