Seal the Chip Case using Chip Sealant:
a.
Hold the syringe tip just above (or in slight contact with the inside wall) of
the fill port of the sealed Chip Case, then carefully fill the port with Chip
Sealant, ensuring that the fluid touches the walls of the port. To ensure the
seal, create a dome of sealant over the top of the port.
IMPORTANT! When applying the Chip Sealant:
·
Apply the Chip Sealant only to the fill port. Do not apply the sealant to
the Chip ID or Chip Case Lid window.
·
When not in use, store the Chip Sealant with the syringe tip attached tip-
first inside the brown plastic bag provided to prevent the sealant in the
syringe from curing.
X0300036
b.
Insert the Digital PCR 20K Chip
assembly into the UV-Curing Station on the
Chip Loader. Push the chip into the station until the ultraviolet light
illuminates.
IMPORTANT! Make sure that the chip is oriented so that the fill port is
directly beneath the beam of ultraviolet light.
IMPORTANT! Do not allow the Digital PCR 20K Chip to contact the roof of
the UV-Curing Station.
X0300036
>15 sec
UV
Chapter 2 Prepare Samples and Load Reactions
Load the chips using the Chip Loader
2
QuantStudio
®
3D Digital PCR System User Guide