5.
Rotate the Chip Loader arm so that the Chip Case Lid solidly contacts the Digital
PCR 20K Chip, firmly press down for 15 seconds to ensure a tight seal, then press
the Lid Nest button and return the Chip Loader arm to its original position.
>15 sec
IMPORTANT! While applying the lid:
·
Press down on the Chip Loader until you reach a hard stop. The arm requires
>20 lbs of force to apply the Chip Case Lid. If you suspect that you are
applying insufficient pressure, increase the duration of the press (>15 seconds).
·
Press the Lid Nest button before you retract the Chip Loader arm.
·
Release the Chip Loader arm gently and do not press down again to avoid
creating a vacuum.
6.
Fill the Chip Case with Immersion Fluid:
a.
Hold the Chip Case by its edges and at a 45° angle so that air can escape
from the loading port as you fill it, then carefully dispense Immersion Fluid
into the port until the
Chip Case contains an air bubble no larger than the fill
port (<2-3 mm in diameter).
After filling the Immersion Fluid, rotate it slightly to reveal any air bubbles
that might be hidden behind the serial number label. If bubbles are present,
add additional Immersion Fluid to remove them.
Note: Minimize air bubble creation when you dispense the fluid; however,
one small air bubble (<2-3 mm in diameter) is acceptable.
X0300036
X0300036
b.
Using a low-lint wipe, remove any excess
Immersion Fluid from the Chip
Case to ensure optimal imaging on the QuantStudio
®
3D Instrument.
Chapter 2 Prepare Samples and Load Reactions
Load the chips using the Chip Loader
2
48
QuantStudio
®
3D Digital PCR System User Guide