c.
Remove the red protective film from a QuantStudio
®
3D Digital PCR Chip
Case Lid, press the Lid Nest button ( ), carefully place the Chip Case Lid
into the Lid Nest as shown ( ), then release the button to clamp the chip in
place.
IMPORTANT! Before applying the Chip Case Lid, make certain that it is
correctly oriented to the Digital PCR 20K Chip assembly.
2
1
d.
Briefly vortex and centrifuge the prepared digital PCR reaction (master mix,
assay, and template), then carefully transfer 14.5 μL of the solution into the
sample-loading port of the Sample Loading Blade
. If the reactions were
placed on ice, allow them to warm prior to loading.
While filling the Sample Loading Blade:
Hold the pipette so that the tip meets the sample-loading port at a
sharp angle and so it does not deflect either blade (forward or rear).
Do not depress the pipette to the second stop. Doing so increases the
chance of introducing an air bubble to the Sample Loading Blade.
Avoid creating air bubbles.
Note: Although the master mix may not distribute evenly in the Sample
Loading Blade immediately after loading, the fluid will spread across the
blade prior to loading.
Chapter 2 Prepare Samples and Load Reactions
Load the chips using the Chip Loader
2
46