Set up the ProFlex
System:
a.
Open the package containing the Chip Adapters and install them.
Note: Each Chip Adapter includes a set of four alignment pegs that fit into
the holes of a flat sample block. The Chip Adapter can fit onto either side of
the sample block (right or left).
CAUTION HOT
b.
Connect the power cord, power on the ProFlex
System, then wait for it to
start (about 45 seconds). Follow the onscreen instructions to complete the
installation.
The ProFlex
System is ready to use when the Main Menu displays.
IMPORTANT! If the ProFlex
System does not start, check the position of
the sample block module. The instrument cannot be powered on if the
module is seated incorrectly or if the lever is not completely closed. If the
ProFlex
System does not start, even after adjusting the sample block
module, or if the screen contains any permanent patterns of lines or bars,
contact technical support for assistance.
c.
Configure the ProFlex
System settings as described in the ProFlex
PCR
System User Guide (Pub. no. MAN0007697).