If you intend to connect your instrument to a network, prepare the physical
network connection:
IMPORTANT! Connecting the instrument to a network is optional and can be
done at a later time if you desire.
Connection Action
Wired Connect one end of a standard category-6 Ethernet cable to the
RJ-45 port on the back panel of the instrument and the other end
to an open network port.
Wireless
1. If required for network setup, record the MAC address of the
SparkLAN WUBR-170GN 802.11n Single-Band Wireless
Dongle, which is located on the underside of the device.
2. Unplug the QuantStudio
®
3D Instrument.
3. Turn the QuantStudio
®
3D Instrument over, then use your
fingers to carefully pry open the rear panel.
IMPORTANT! The rear panel is secured to the instrument
by three pegs that require significant force to decouple. If
necessary, insert a flathead screwdriver into the slot at the
base of the panel to pry it from the instrument chassis.
4. Turn the instrument over, then uncap the wireless dongle and
insert it into the USB port
within
the instrument. When the
device is seated within the USB port, firmly push on the
wireless dongle to ensure that it is plugged in completely.
Note: The internal USB port is accessed through a small
opening in the instrument chassis.
5. Reattach the rear panel to the instrument chassis.
Chapter 1 Getting Started
Prepare the QuantStudio
®
3D Digital PCR System for use
1
QuantStudio
®
3D Digital PCR System User Guide