When planning the placement of the QuantStudio
®
3D Digital PCR System
components, give special consideration to the design and organization of the
laboratory. To minimize the potential for environmental contamination while
preparing your dPCR reactions, we recommend organizing your laboratory so that
the area where amplified DNA is handled is physically isolated from the work areas
for DNA extraction and PCR setup. These pre- and post-PCR designated work areas,
each ideally with dedicated equipment and supplies, will allow you to maintain strict
physical isolation between the area designated for handling amplified DNA and the
area where nucleic acids are extracted and dPCR reactions are assembled, thereby
avoiding transfer of amplified DNA out of the designated work area.
Room Activities Equipment
PCR Setup Work
Area (pre-PCR
room)
Sample preparation and DNA
extraction
Extracted DNA sample storage
PCR reaction preparation and
DNA sample additions
QuantStudio
®
3D Digital PCR 20K
Chip loading and sealing
General laboratory supplies (such as gloves and
wipes)
Microcentrifuge
Vortexer
Pipettes, P10 to P1000 with tips
QuantStudio
®
3D Digital PCR 20K Chip
QuantStudio
®
3D Digital PCR Chip Case Lid
QuantStudio
®
3D Digital PCR Chip Loader
QuantStudio
®
3D Digital PCR Sample Loading Blade
Immersion Fluid
UV-Activated Chip Sealant Syringes and Tips
Amplified DNA
Work Area (post-
PCR room)
PCR amplification
QuantStudio
®
3D Digital PCR 20K
Chip reading and storage
Sealed QuantStudio
®
3D Digital
PCR 20K Chip storage
ProFlex
2x Flat PCR System (or Dual Flat Block
GeneAmp
®
PCR System 9700)
QuantStudio
®
3D Digital PCR Instrument
Note: Ideally, amplified DNA or equipment and supplies used to handle amplified
DNA should not be taken out of the designated work area. If the work area for
amplified DNA is in a separate but contiguous room, make sure that air flows toward
the amplified DNA area. In addition, it is helpful if there is a separate exit from the
post-PCR work area that does not exit into the pre-PCR work area.
Note: If possible, the station for loading Digital PCR 20K Chips should be physically
separated from the PCR area, away from the ProFlex
System (or GeneAmp
®
PCR
System 9700) and the QuantStudio
®
3D Instrument.